AccuPen手持压平式眼压计测量的3岁以下婴幼儿正常眼压值研究

苏晓丹  赵书萍  邓敏    

  1. 650100昆明儿童医院眼科
  • 收稿日期:2018-05-11 出版日期:2019-01-25 发布日期:2019-01-29
  • 通讯作者: 苏晓丹,Email:xiaodannihao2007@163.com
  • 基金资助:

    云南省教育厅科学研究基金项目(2017zDX180)

Intraocular pressure with AccuPen tonometer in children aged 3 or younger

SU Xiao-dan, ZHAO Shu-ping, DENG Min.   

  1. Department of Ophthalmology, Kunming Children's Hospital, Kunming 650100, China
  • Received:2018-05-11 Online:2019-01-25 Published:2019-01-29
  • Contact: SU Xiao-dan, Email: xiaodannihao2007@163.com

摘要:

目的 研究0~3岁婴幼儿正常眼压值及其影响因素。设计 横断面调查。研究对象 0~3岁正常婴幼儿514例。方法 水合氯醛镇静下使用AccuPen手持数字化压平式眼压计对每只眼测量仰卧位眼压3次。并使用手持角膜测厚仪测量中央角膜厚度。眼压与角膜厚度进行Pearson 相关分析。主要指标 眼压值。结果 514例中右眼平均眼压(15.63±2.76)mmHg,左眼(16.17±2.66)mmHg。0~12个月组454例,右眼平均眼压(15.69±2.82)mmHg,左眼(16.26±2.69)mmHg;13~24个月组51例,右眼平均眼压(15.31±2.16)mmHg,左眼(15.51±2.19)mmHg;25~36个月组9例,右眼平均眼压(14.67±2.92)mmHg,左眼(15.11±2.85)mmHg。眼压与中央角膜厚度呈弱正相关(r=0.302~0.313,P=0.000)。结论 0~3岁婴幼儿眼压值与成年人无明显差别。(眼科,2019, 28: 49-51)

关键词: 婴幼儿, 眼压

Abstract:

 Objective To study the normal intraocular pressure (IOP) and its influencing factors in infants aged 0-3 years. Design Cross-sectional survey. Participants 514 cases of normal infants aged 0-3 years. Methods Under the sedation of chloral hydrate, IOP in supine position was measured three times in each eye with AccuPen hand-held digital tonometer. The central corneal thickness was measured by hand-held corneal thickness gauge. Pearson correlation analysis was performed between IOP and corneal thickness. Main Outcome Measures IOP. Results The mean IOP in the right eye was (15.63±2.76) mmHg, and that in the left eye was (16.17±2.66) mmHg. There were 454 cases in 0-12 months group, mean IOP in right eye (15.69±2.82) mmHg), left eye (16.26±2.69) mmHg; 51 cases in 13-24 months group, mean IOP in right eye (15.31±2.16) mmHg, left eye (15.51±2.19) mmHg; 9 cases in 25-36 months group, mean IOP in right eye (14.67±2.92) mmHg and left eye (15.11±2.85) mmHg. There was a weak positive correlation between IOP and central corneal thickness (r=0.302-0.313, P=0.000). Conclusion There is no significant difference in IOP between 0-3 year’s old infants and adults. (Ophthalmol CHN, 2019, 28: 49-51)

Key words:  infants; , intraocular pressure