圆锥角膜的光密度分析及其与角膜形态学的相关性研究

张梦雨1   田磊1,2   郭莉莉1   张阳1   王智群1   陈可欣1   

  1. 1首都医科大学附属北京同仁医院  北京同仁眼科中心 眼科学与视觉科学北京市重点实验室 北京市眼科研究所 100730; 2北京大数据精准医疗高精尖创新中心  北京航空航天大学&首都医科大学,北京同仁医院 100730
  • 收稿日期:2020-04-02 出版日期:2020-07-22 发布日期:2020-07-21
  • 通讯作者: 田磊,Email:tianlei0131@163.com
  • 基金资助:
    国家自然科学基金资助项目(31600758);北京市科技新星计划(Z181100006218099);北航-首医大数据精准医疗高精尖创新中心同仁分中心开放基金项目(BHTR-KFJJ-202001)

Corneal densitometry analysis and its correlation with corneal topography in keratoconus

Zhang Mengyu1, Tian Lei1,2, Guo Lili1, Zhang Yang1, Wang Zhiqun1, Chen Kexin1   

  1. 1 Beijing Tongren Eye Center, Beijing Key Laboratory of Ophthalmology and Visual Sciences, Beijing Institute of Ophthalmology, Beijing Tongren Hospital, Capital Medical University; Beijing 100730, China; 2 Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beihang University & Capital Medical University, Beijing Tongren Hospital, Beijing 100730, China
  • Received:2020-04-02 Online:2020-07-22 Published:2020-07-21
  • Contact: Tian Lei, Email: tianlei0131@163.com
  • Supported by:
    National Natural Science Foundation of China (31600758); Beijing Nova Program (Z181100006218099); the Open Research Fund from Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beijing Tongren Hospital, Beihang University & Capital Medical University (BHTR-KFJJ-202001)

摘要:  目的 对比正常角膜与圆锥角膜(KC)的角膜光密度(CD),并分析CD与角膜形态学参数之间的相关性。设计 横断面研究。研究对象 2017-2019年北京同仁医院KC患者162例(162眼)(KC组)和健康志愿者41例(41眼)(NL组)。 方法 用Pentacam三维眼前节分析系统测量CD,以角膜顶点为中心,分别测量0~2 mm、2~6 mm、6~10 mm、10~12 mm和0~12 mm直径范围的前层(120 μm)、中间层、后层(60 μm)以及全层的CD。基于Pentacam的TKC分级法将KC按照严重程度分为轻度KC(TKC=1)、中度KC(TKC=1~2,2)和重度KC(TKC=3,3~4,4)。Bonferroni矫正检验或Kruskal-Wallis检验比较组间CD差异。Pearson相关分析或Spearman相关分析评估CD与角膜形态学参数之间的相关性。主要指标 不同直径范围和深度层面的CD、角膜曲率最大值(Kmax)、中央角膜厚度(CCT)、角膜最薄点厚度(TCT)。 结果 NL组,总KC组,轻度、中度及重度KC组在0~2 mm前层的CD分别为21.67±1.60,21.90(18.70~98.70),21.6(18.70~24.80),21.60(19.30~29.10)及22.80(20.00~98.70)。对于0~2 mm前层的CD,总KC组较NL组显著增高(P=0.043),轻度、中度KC组与NL组之间无显著差异(P均>0.05),而重度KC组较轻度、中度KC组和NL组均显著增高(P均<0.05)。其他区域CD组间均无统计学差异(P均>0.05)。KC组0~2 mm前层的CD与Kmax显著正相关(R=0.181,P=0.03),与CCT和TCT显著负相关(R=-0.172,P=0.03;R=-0.171,P=0.04)。结论 KC前层角膜中央区的CD较正常人高,且KC越严重CD越高。KC前层角膜中央区的CD与Kmax、CCT和TCT显著相关。CD或许可作为评估KC进展程度的指标之一。(眼科,2020,29: 278-283)

关键词: 角膜光密度, 圆锥角膜, 角膜最大曲率, 角膜厚度

Abstract: Objective To compare corneal densitometry(CD) between keratoconus(KC) and normal cornea, and to investigate the correlation between CD and topographic parameters. Design Cross-sectional study. Participants A total of 162 keratoconic patients (162 eyes) (KC group) and 41 normal subjects (41 eyes) (NL group) were included from Beijing Tongren Hospital from 2017 to 2019. Methods CD values were obtained for the 0~2 mm, 2~6 mm, 6~10 mm, 10~12 mm and 0~12 mm zones of the anterior (up to 120 μm), posterior (posterior 60 μm), central (between the anterior and posterior) and total cornea by Pentacam HR. The Pentacam provides a topographic keratoconus classification for staging the disease in grades 1 to 4. Cases were divided into mild, grade 1; moderate, from grades 1 to 2 and grade 2; and severe, grade 3, from grades 3 to 4 and grade 4. Bonferroni multiple comparisons test or Kruskal-Wallis test was used to compare CD between different groups. Pearson’s correlation test or Spearman’s correlation test was used to determine the correlations between CD and topographic parameters. Main Outcome Measures CD values over different corneal annular concentric zones and depths, maximum keratometry (Kmax), corneal central thickness (CCT) and thinnest corneal thickness (TCT). Results The CD values of the anterior 0 to 2 mm layer for the NL group, KC group, KC mild, KC moderate and KC severe were 21.67±1.60, 21.90(18.70~98.70), 21.6(18.70~24.80), 21.60(19.30~29.10) and 22.80(20.00~98.70), respectively. There were significant differences between KC group and NL group for the CD of the anterior 0 to 2 mm layer (P=0.043). There were no significant differences between KC mild, KC moderate and NL group for the CD of the anterior 0 to 2 mm layer(all P>0.05). There were significant differences between KC severe and the other groups for the CD of the anterior 0 to 2 mm layer (all P<0.05). There were no significant differences between groups for the CD of the remaining corneal annuluses and layers(all P>0.05). The CD values of the anterior 0 to 2 mm layer significantly and positively correlated with the Kmax values (R=0.181,P=0.03), and significantly and negatively correlated with CCT and TCT values (R=-0.172, P=0.03 and R=-0.171, P=0.04, respectively). Conclusion The CD values of the anterior 0 to 2 mm layer are higher for the keratoconic cornea than in the normal cornea. The CD level is higher in more advanced stages. There are significant correlations between the CD values of the anterior 0 to 2 mm layer of the keratoconic cornea and the Kmax, CCT and TCT values. CD may be a useful index of KC progression. (Ophthalmol CHN, 2020, 29: 278-283)

Key words: corneal densitometry, keratoconus, maximum keratometry, corneal thickness