眼科 ›› 2025, Vol. 34 ›› Issue (2): 146-151.doi: 10.13281/j.cnki.issn.1004-4469.2025.02.010

• 论著 • 上一篇    下一篇

偏心对不同屈光度的景深延长型人工晶状体光学质量影响的实验研究

秦苏云1   廖萱  兰长骏  谢丽暄  潘若琳1   谭青青1   黄欢王艳1   

  1. 1川北医学院附属医院眼科  川北医学院眼视光医学院,四川南充637000; 2 成都东区爱尔眼科医院 成都高新区爱尔眼科研究所,成都610056
  • 收稿日期:2024-10-25 出版日期:2025-03-25 发布日期:2025-03-13
  • 通讯作者: 兰长骏,Email: eyelanchangjun@163.com
  • 基金资助:
     四川省医学科研课题(S21011);成都市医学科研课题(2024268);爱尔眼科医院集团临床研究所科研基金项目(AIG2306D01)

Experimental study on the influence of decentration on the optical properties of extended depth of focus intraocular lenses with different diopter

Qin Suyun1, Liao Xuan1, Lan Changjun2, Xie Lixuan1, Pan Ruolin1, Tan Qingqing1, Huang Huan1, Wang Yan1   

  1. 1 Department of Ophthalmology, Affiliated Hospital of North Sichuan Medical College; Medical School of Ophthalmology & Optometry, North Sichuan Medical College, Nanchong Sichuan 637000, China; 2 Chengdu Eastern Aier Eye Hospital; AIER Eye Institute in Chengdu High-Tech Zone, Chengdu 610056, China
  • Received:2024-10-25 Online:2025-03-25 Published:2025-03-13
  • Contact: Lan Changjun, Email: eyelanchangjun@163.com
  • Supported by:
    The Medical Research Project of Sichuan Province (S21011); The Medical Research Project of Chengdu City (2024268); Clinic Research Foundation of Aier Eye Hospital Group (AIG2306D01)

摘要: 目的  探讨偏心对不同屈光度景深延长型人工晶状体(EDOF IOL)光学质量的影响。设计  实验研究。研究对象  不同屈光度TECNIS Symfony ZXR00 EDOF IOL。方法  采用体外光学质量测试系统OptiSpheric IOL R&D评价+10 D、+20 D、+30 D的ZXR00,在 ISO-2模型角膜下获取IOL的离焦曲线和离焦美国空军(USAF)分辨率测试图,测定IOL处于居中位置及偏心0.3、0.5、0.7、0.9及1.1 mm时焦点处50 lp/mm的调制传递函数(MTF)值、MTF曲线和USAF分辨率测试图。主要指标  MTF值、MTF曲线和USAF分辨率测试图。结果  USAF分辨率测试图显示离焦0~2.00 D范围ZXR00具有良好且稳定的图像质量。居中位置远焦点处+10 D、+20 D、+30 D ZXR00的50 lp/mm MTF值分别为0.391、0.410、0.456,中焦点处的MTF值分别为0.321、0.381、0.362。偏心0.3 mm时3种屈光度IOL的MTF值开始下降,偏心0.5 mm时MTF值继续下降,USAF图像显示偏心0.5 mm内3种屈光度IOL尚保持相对较好的图像质量。远焦点处偏心0.7 mm时+10 D、+20 D、+30 D ZXR00的MTF值分别为0.228、0.255、0.332,USAF图像显示较显著的图像质量下降。MTF曲线显示远焦点处高屈光度IOL光学质量降幅更小,中焦点处低屈光度IOL降幅更小。发生偏心后MTF曲线在高频区降幅更大,低频区降幅较小,25 lp/mm空间频率低屈光度IOL受到的影响更小。结论  EDOF IOL ZXR00具有较好且连续的远中距离成像质量。偏心对IOL的光学质量产生负面影响,偏心0.5 mm及以内,3种屈光度IOL均有较好的抗偏心能力。不同屈光度IOL受到的负面影响程度不一,远、中焦点受到的影响也有差别。(眼科,2025,34: 146-151)

关键词: 人工晶状体, 景深延长, 体外测试, 偏心

Abstract: Objective To evaluate the impact of decentration on the optical quality of EDOF IOL with different diopters. Design Experimental study. Participants TECNIS Symfony ZXR00 EDOF IOL with different diopters. Methods OptiSpheric IOL R&D was used to evaluate the optical quality of the +10 D, +20 D, and +30 D ZXR00. Under the ISO-2 model cornea conditions, the defocus curves and defocus USAF images were obtained. The modulation transfer function (MTF) values at spatial frequencies of 50 lp/mm, MTF curves, and USAF resolution test charts were measured at each focal point of the IOL when the IOL was in the centered position and at decentration of 0.3, 0.5, 0.7, 0.9, and 1.1 mm. Main Outcome Measures MTF values, MTF curves, and USAF resolution test charts. Results The USAF resolution test charts demonstrated that the ZXR00 maintained good and stable image quality within a defocus range of 0 D to 2.00 D. At the centered position, the 50  lp/mm MTF values for the +10 D, +20 D, and +30 D ZXR00 at the far focus were 0.391, 0.410, and 0.456, respectively, while at the intermediate focus, the values were 0.321, 0.381, and 0.362, respectively. When decentered by 0.3 mm, the MTF values of three types of diopter IOLs began to decline. And this decline continued at 0.5 mm decentration. The USAF images showed that three types of diopter IOLs maintained relatively good image quality within a decentration of 0.5 mm. With a decentration of 0.7 mm, the MTF values for the +10 D, +20 D, and +30 D IOLs at the far focus were 0.228, 0.255, and 0.332, respectively, with the USAF images showing a significant decline in image quality. The MTF curves revealed that IOLs with higher diopters experienced a smaller decline in optical quality at the far focus, while IOLs with lower diopters showed a smaller decline at the intermediate focus. After decentration, the MTF curves exhibited a greater decline in the high-frequency range and a smaller decline in the low-frequency range. At the spatial frequency of 25 lp/mm, lower-diopter IOLs were less affected. Conclusion The EDOF IOL ZXR00 demonstrated consistent and stable image quality from far to intermediate distances. Decentration had a negative impact on the optical quality of the IOLs. Within a decentration of 0.5 mm, the three types of diopter IOLs showed good resistance to decentration. The degree of negative impact varied among different diopter IOLs, and the effects on the far and intermediate focus also differed. (Ophthalmol CHN, 2025, 34: 146-151)

Key words:  intraocular lens, extended depth-of-focus, in vitro test, decentration