眼科 ›› 2025, Vol. 34 ›› Issue (4): 279-284.doi: 10.13281/j.cnki.issn.1004-4469.2025.04.005

• 论著 • 上一篇    下一篇

景深延长型和区域折射型人工晶状体植入术后视觉质量比较

万雨  宋旭东  陶靖   

  1. 首都医科大学附属北京同仁医院 北京同仁眼科中心 眼科学与视觉科学北京市重点实验室,北京 100730
  • 收稿日期:2024-12-24 出版日期:2025-07-25 发布日期:2025-07-13
  • 通讯作者: 陶靖,Email:taojing@mail.ccmu.edu.cn
  • 基金资助:
    北京市科技新星计划资助(20240484601);首都医科大学附属北京同仁医院科研种子基金(2021-YJJ-ZZL-046)

Comparison of visual quality after implantation of extended depth-of-focus and rotational asymmetric refractive intraocular lenses

Wan Yu, Song Xudong, Tao Jing   

  1. Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University; Beijing Key Laboratory of Ophthalmology and Visual Sciences, Beijing 100730, China
  • Received:2024-12-24 Online:2025-07-25 Published:2025-07-13
  • Contact: Tao Jing, Email: taojing@mail.ccmu.edu.cn
  • Supported by:
     Sponsored by Beijing Nova Program (20240484601); The Priming Scientif Research Foundation for the Junior Researcher in Beijing Tongren Hospital, Capital Medial University (2021-YJJ-ZZL-046)

摘要:  目的 比较景深延长型和区域折射型人工晶状体(IOL)植入术后的视觉质量。设计  回顾性队列研究。研究对象  2023年7月至2024年10月北京同仁医院接受白内障超声乳化吸除联合IOL植入术的患者78例(110眼)。方法  根据植入IOL的不同,分为景深延长组(植入TECNIS Symfony ZXR00 IOL)46例(62眼)和区域折射组(植入SBL-3 IOL)32例(48眼)。在术后3个月时比较两组患者远、中、近视力,离焦曲线及术后视功能满意度问卷(NEI VFQ-25)调查结果,并通过OPD-ScanⅢ屈光分析仪检测客观视觉质量。主要指标  术后远、中、近视力,离焦曲线,视功能满意度评分,全眼和高阶像差,斯特雷尔比(SR)及调制传递函数(MTF)。结果  术后3个月两组裸眼远、近视力无显著性差异,景深延长组裸眼中视力为0(0,0.10)Log MAR,显著优于区域折射组的0.10(0,0.20)Log MAR(P<0.001)。两组术后离焦曲线显示,景深延长组在0.0~-2.5 D区间内呈平缓走势,区域折射组在0.0 D和-2.5 D处呈双峰状。两组问卷调查在总分及各维度评分无显著性差异。术后3个月,3 mm瞳孔直径下景深延长组球差为0.01(0,0.02)μm,小于区域折射组的0.02(0.01,0.03)μm(P=0.017),4 mm时两组球差无显著性差异,5、6 mm时景深延长组球差大于区域折射组(P=0.020,0.030);3~5 mm瞳孔直径时景深延长组彗差均小于区域折射组(P均<0.001),6 mm时两组彗差无显著性差异;景深延长组在3~6 mm瞳孔直径下的三叶草像差、全眼像差均小于区域折射组(P均<0.001);术后3个月,景深延长组的SR、MTF均高于区域折射组(P均<0.001)。结论  景深延长型和区域折射型IOL植入术后均能提供有效全程视力和良好视觉质量,前者中视力更优,但其球差易受瞳孔大小影响。

关键词:  , 景深延长型人工晶状体;区域折射型人工晶状体;视觉质量

Abstract:  Objective  To compare the visual quality after implantation of extended depth-of-focus (EDOF) and rotational asymmetric refractive intraocular lenses(IOL). Design  Retrospective cohort study. Participants  A total of 78 patients (110 eyes) who underwent phacoemulsification combined with IOL implantation at Beijing Tongren Hospital between July 2023 and October 2024. Methods  Based on the type of IOL implanted, patients were divided into the EDOF group (TECNIS Symfony ZXR00 IOL) included 46 patients (62 eyes), and rotational asymmetric refractive group (SBL-3 IOL) comprised 32 patients (48 eyes). At 3 months postoperatively, uncorrected distance (UDVA), intermediate (UIVA), and near (UNVA) visual acuity, defocus curves, and visual function satisfaction questionnaire (NEI VFQ-25) results were compared. Objective visual quality was assessed using the OPD-Scan III aberrometer. Main Outcome Measures  Postoperative distance, intermediate, and near visual acuity; defocus curves; visual satisfaction scores; total and higher-order aberrations (HOAs); Strehl ratio (SR); and modulation transfer function (MTF). Results  At 3 months postoperatively, no significant differences were observed in UDVA or UNVA between the two groups. The EDOF group demonstrated significantly better in UIVA than the rotational asymmetric refractive group (P<0.001). Defocus curves revealed a flat profile from 0.0 to -2.5 D in the EDOF group, whereas the rotational asymmetric refractive group exhibited a bifocal pattern with peaks at 0.0 D and -2.5 D. No significant differences were found in NEI VFQ-25 total or subscale scores. At 3 mm pupil diameter, the EDOF group showed lower spherical aberration than the rotational asymmetric refractive group (P=0.017). At 5 mm and 6 mm pupil diameters, spherical aberration was higher in the EDOF group (P=0.020, 0.030). Coma aberration was lower in the EDOF group at 3~5 mm pupil diameters (all P<0.001). Trefoil and total ocular aberrations were significantly lower in the EDOF group across all pupil diameters (all P<0.001). The EDOF group also exhibited higher SR and MTF values compared to the rotational asymmetric refractive group (all P<0.001). Conclusion  Both EDOF and rotational asymmetric refractive IOL provide effective full-range vision and satisfactory visual quality. The EDOF IOL offers superior intermediate vision, though its spherical aberration is more influenced by pupil size.

Key words:  Extended depth-of-focus IOL, Rotational asymmetric refractive IOL, Visual quality